3 monopolar cutting modes: pure cut, blend 1, blend 2, blend 3
Pure cut: cut the tissue cleanly and accurately without coagulation.
blend 1: Use when the cutting speed is slightly slow and a small amount of hemostasis is required.
blend 2: Compared with blend 1, it is used when the cutting speed is slightly slower and a better hemostatic effect is needed.
2 coagulation modes: spray coagulation, forced coagulation, and soft coagulation
forced coagulation: It’s non-contact coagulation. The output threshold voltage is lower than spray coagulation. It’s suitable for coagulation in a small area.
spray coagulation: high-efficiency coagulation without contact surface. The coagulation depth is shallow. The tissue is removed by evaporation. It usually uses a Blade or ball electrode for coagulation.
1 bipolar output mode: Vessel Sealing mode:
Provide exceptional coagulation and transection of vessels up to 7mm in diameter.
Mode |
Max Output Power(W) |
Load impedance (Ω) |
Modulation Frequency (kHz) |
Max Output Voltage (V) |
Crest Factor |
||
Monopolar |
Cut |
Pure Cut |
100 |
500 |
—— |
1300 |
1.8 |
Blend 1 |
100 |
500 |
20 |
1400 |
2.0 |
||
Blend 2 |
100 |
500 |
20 |
1300 |
2.0 |
||
Coag |
Spray |
90 |
500 |
12-24 |
4800 |
6.3 |
|
Forced |
60 |
500 |
25 |
4800 |
6.2 |
||
Bipolar |
Vessel sealing |
60 |
100 |
20 |
700 |
1.9 |
Product Name |
Product Number |
Monopolar Foot-Switch | JBW-200 |
Hand-Switch Pencil, Disposable | HX-(B1)S |
Patient Return Electrode Rods (10mm) With Cable, Reusable | 38813 |
5mm ,37cm Length Laparoscopic Instrument Straight Tip | SM1150 |